This article is from Surgical Neurology International , volume 3 . Abstract Background:: Exposure of externally programmable shunt-valves (EPS-valves) to magnetic resonance imaging (MRI) may lead to unexpected changes in shunt settings, or affect the ability to reprogram the valve. We undertook this study to examine the effect of exposure to a 3T MRI on a group of widely used EPS-valves. Methods:: Evaluations were performed on first generation EPS-valves (those without a locking mechanism to prevent changes in shunt settings by external magnets other than the programmer) and second generation EPS-valves (those with a locking mechanisms). Fifteen new shunt-valves were divided into five groups of three identical valves each, and then exposed to a series of six simulated MRI scans. After each of the exposures, the valves were evaluated to determine if the valve settings had changed, and whether the valves could be reprogrammed. The study produced 18 evaluations for each line of shunt-valves. Results:: Exposure of the first generation EPS-valves to a 3T magnetic field resulted in frequent changes in the valve settings; however, all valves retained their ability to be reprogrammed. Repe