This article is from Indian Journal of Pharmaceutical Sciences , volume 74 . Abstract A gradient reversed phase HPLC method was developed and validated for analysis of betamethasone dipropionate, its related substances and degradation products, using Altima C18 column (250×4.6 mm, 5 μm) with a flow rate of 1.0 ml/min and detection wavelength of 240 nm. The mobile phase A is a mixture of water, tetrahydrofuran and acetonitrile in the ratio of 90:4:6 (v/v/v) while mobile phase B is a mixture of acetonitrile, tetrahydrofuran, water and methanol in the ratio of 74:2:4:20 (v/v/v/v). The samples were analyzed using 20 μl injection volume and the column temperature was maintained at 50°. The limit of detection and limit of quantitation were found to be 0.02 μg/ml and 0.07 μg/ml, respectively. The stability-indicating capability of method was established by forced degradation studies and method demonstrated successful separation of drug, its related substances and degradation products. The method was validated as per the International Conference on Harmonization guidelines. The developed method is linear in the range of 0.07 to 200% of specification limits established for all the known rel