This article is from Annals of Indian Academy of Neurology , volume 16 . Abstract Objective:: The objective of this study is to assess the safety, feasibility and clinical value of transcranial motor evoked potential (MEP) monitoring by electrical stimulation. Setting:: Clinical neurophysiology department of tertiary reach hospital. Materials and Methods:: MEP monitoring was attempted in 44 “high risk” patients. Intraoperative surgical, anesthesia and neurophysiological findings were documented prospectively. MEP monitoring results were correlated with motor outcome. Results:: The success for reliable MEP recording from the lower limbs was 75%. Incidence of new permanent post-operative motor deficit was zero. Nearly, 76.5% of the cases (13 out of 17 cases) who showed unobtainable and unstable MEP outcome had lesion location in the spine as compared with 23.5% (4 out of 17 cases) that had lesion location in the brain. Chi-square test demonstrated a statistically significant difference between these two groups (P = 0.0020). Out of these 13 spine surgery cases, 8 (62%) were operated for deformity. Seven out of 12 (60%) patients less than 12 years of age had a poor MEP monitoring outco