This article is from Dental Research Journal , volume 11 . Abstract Background:: The aim of this in vitro study was to evaluate the effect of shape, diameter and length of implants on their primary stability based on resonance frequency analysis. Materials and Methods:: Replace select tapered and Branemark MK III implants were selected. Each of these two selected groups was divided into nine subgroups based on the implant length (IL) (short, medium and long) and the implant diameter (ID) (narrow platform [NP], regular platform [RP] and wide platform [WP]). Five implants were assigned to each of the nine subgroups. Implants were placed in artificial bone blocks with bone quality similar to D3 bone. Immediately after the implant placement, its primary stability was measured using Osstell Mentor equipment. T-test and Tukey's honest significant difference Post hoc were performed for data analysis. Statistical significance was defined at P < 0.05. Results:: Replace select system showed significantly higher primary stability compared to the Branemark system, when using the short implants for all three diameters (P ≤ 0.004). However, in medium length implants there were no significant dif