This article is from Radiotherapy and Oncology , volume 110 . Abstract Background and purpose: To develop a class solution for prostate Stereotactic Ablative Radiotherapy (SABR) using Volumetric Modulated Arc Therapy (VMAT). Materials and methods: Seven datasets were used to compare plans using one 360° arc (1FA), one 210° arc (1PA), two full arcs and two partial arcs. Subsequently using 1PA, fifteen datasets were compared using (i) 6 mm CTV–PTV margins, (ii) 8 mm CTV–PTV margins and (iii) including the proximal SV within the CTV. Monaco™ 3.2 (Elekta™) was used for planning with the Agility™ MLC system (Elekta™). Results: Highly conformal plans were produced using all four arc arrangements. Compared to 1FA, 1PA resulted in significantly reduced rectal doses, and monitor units and estimated delivery times were reduced in six of seven cases. Using 6 mm CTV–PTV margins, planning constraints were met for all fifteen datasets. Using 8 mm margins required relaxation of the uppermost bladder constraint in three cases to achieve adequate coverage, and, compared to 6 mm margins, rectal and bladder doses significantly increased. Including the proximal SV required relaxation of the uppermost
Sign in with your free account to read, download EPUBs or listen.
Ebook · Developing a class solution for Prostate Stereotactic Ablative Body Radiotherapy (SABR) using Volumetric Modulated Arc Therapy (VMAT). · Murray, Louise J., Cosgrove, Vivian, Lilley, John, Sykes, Jonathan, Thompson, Christopher M., Franks, Kevin, Sebag-Montefiore, David, Henry, Ann M. · Internet Archive