English
This article is from Surgical Neurology International , volume 5 . Abstract Background:: Patient-specific implants are used for cranioplastic skull reconstruction when large bone flaps must be replaced or where there are complex or critical contours, especially near the face. These implants have a low complication rate, with poor fit and postoperative infection being the most common complications. We report here a potentially serious hazard that may arise from the use of porous implants. Case Description:: A 45-year-old woman sustained severe head trauma in a motor vehicle accident that required urgent surgical intervention. Because of progressive resorption of her native bone flap, she underwent replacement of her native flap with a hard tissue replacement/patient-matched implant cranioplasty. Eight years later, she sustained a traumatic laceration over her vertex that necessitated removal of her cranioplastic implant because of persistent local infection. Intraoperatively, the dural flap was ingrowing and firmly adherent to the inside surface of the porous cranioplasty. After several failed attempts to remove the whole implant piecemeal, we attempted to dissect the dural flap fro
Sign in with your free account to read, download EPUBs or listen.
Ebook · Dural adhesion to porous cranioplastic implant: A potential safety concern. · Sayama, Christina M., Sorour, Mohammad, Schmidt, Richard H. · Internet Archive