This article is from Journal of Medical Physics / Association of Medical Physicists of India , volume 39 . Abstract An automated Multi-Leaf Collimator (MLC) system has been developed as add-on for the cobalt-60 teletherapy machines available in India. The goal of the present computational study is to validate the MLC design using Monte Carlo (MC) modeling. The study was based on the Kirloskar-supplied Phoenix model machines that closely match the Atomic Energy of Canada Limited (AECL) theratron-80 machine. The MLC is a retrofit attachment to the collimator assembly, with 14 non-divergent leaf pairs of 40 mm thick, 7 mm wide, and 150 mm long tungsten alloy plates with rounded edges and 20 mm tongue and 2 mm groove in each leaf. In the present work, the source and collimator geometry has been investigated in detail to arrive at a model that best represents the measured dosimetric data. The authors have studied in detail the proto-I MLC built for cobalt-60. The MLC field sizes were MC simulated for 2 × 2 cm2 to 14 × 14 cm2 square fields as well as irregular fields, and the percent depth dose (PDD) and profile data were compared with ROPS† treatment planning system (TPS). In addition,