This article is from Journal of Clinical Imaging Science , volume 4 . Abstract Objectives:: This study compares cancerogenesis risks posed by the 64 row detector and the 320 row detector computed tomography scanners used during coronary computed tomography angiography (CCTA) following decennial screening guidelines. Material and Methods:: Data of the radiation absorbed after CCTA by lung, thyroid, and female breast in patients between 50 and 70 years of age obtained from prior published literature for the 64 row CT scanner were compared with data from our study using 320 row detector CT scanner. Data from the 64 row and the 320 row detector CT scanners was used to determine lifetime attributable risks (LAR) of cancer based on the biological effects of ionizing radiation (BEIR) VII report. Results:: The relative reduction of LAR (%) for 50-, 60-, and 70-year-old patients undergoing scanning with the 320 row detector CT scanner was 30% lower for lung, and more than 50% lower for female breast when compared with results from 64 row detector CT scanner. The use of 320 row detector CT would result in a combined cumulative cancer incidence of less than 1/500 for breast in women and less