English
This article is from Journal of Orthodontic Science , volume 1 . Abstract Objectives:: To investigate the reliability of mechanical and adhesive methods of fixing rectangular wires in the cross-slot of a mini-implant. Materials and Methods:: A twin-hooked wire element was placed and fixed in the slot of a mini-implant via a NiTi spring under tension, or by means of an adhesive. For the purpose of mechanical anchorage, the wire was crimped with a special crimping tool to increase its thickness and prevent it from slipping through the slot. Before applying the adhesive, there were four possible methods of preparation: Untreated Wire (Adh. 1); ROCATEC-Pre (Adh. 2); ROCATEC-PRE + Espe Sil (Adh. 3); ROCATEC-PRE + ROCATEC-PLUS + Espe Sil (Adh. 4). The mechanical fixing and two adhesive fixings were aged by means of temperature change (500 cycles). A Zwick universal testing machine was used to measure the maximum strength of the coupling. Results:: In all tests, the untreated wire + adhesive withstood the lowest maximum load (Ø 3 N and 10.8 N respectively) and failed the aging test after a maximum of eight cycles. The wires in test group Adh. 4 withstood the highest maximum load (Ø 43.3 N
Sign in with your free account to read, download EPUBs or listen.
Ebook · Reliable coupling of orthodontic elements to mini-implants: An in-vitro study. · Ludwig, Bjorn, Zipprich, Holger, Gebel, Britta, Piller, Ben, Antoszewska, Joanna · Internet Archive