English
This article is from Journal of Medical Physics / Association of Medical Physicists of India , volume 39 . Abstract The dosimetric accuracies of CATPhan 504 and CIRS 062 have been evaluated using the kV-CBCT of Varian TrueBeam linac and Eclipse TPS. The assessment was done using the kV-CBCT as a standalone tool for dosimetric calculations towards Adaptive replanning. Dosimetric calculations were made without altering the HU-ED curves of the planning computed tomography (CT) scanner that is used by the Eclipse TPS. All computations were done using the images and dataset from kV-CBCT while maintaining the HU-ED calibration curve of the planning CT (pCT), assuming pCT was used for the initial treatment plan. Results showed that the CIRS phantom produces doses within ±5% of the CT-based plan while CATPhan 504 produces a variation of ±14% of the CT-based plan.
Sign in with your free account to read, download EPUBs or listen.
Ebook · Assessment of the dosimetric accuracies of CATPhan 504 and CIRS 062 using kV-CBCT for performing direct calculations. · Annkah, James Kwame, Rosenberg, Ivan, Hindocha, Naina, Moinuddin, Syed Ali, Ricketts, Kate, Adeyemi, Abiodun, Royle, Gary · Internet Archive